Reproducibility
Overview
The Reproducibility functionality allows you to evaluate the deviation of the peaks height/area assigned to a substance. In other words, the reproducibility of the measurement between different tracks is computed.
During the definition
In the Definition tab of the evaluation, use the Advanced options button to make the Reproducibility field appear. This field can be activated/deactivated independently for each substance. In this mode:
the integration and substance assignment phases are done the same way as for other substances,
no calibration is performed, that’s why the Calibration mode and Range deviation fields are hidden, and
as a result, there’s no difference between sample and reference tracks.
In the results
In the Results tab, the reproducibility results are displayed in a separate tab. For each substance, the coefficient of variation (main result of the reproducibility measurement) and the average height/area are displayed. A grid view recalls the height or the area of the peaks assigned to the substance. Additionally, a graphic is displayed to help the visualization of the deviation of each value from the average.
The deviation from the average is expressed in percent.
In the graphic view, a red line shows the average value (deviation = 0%).
It’s possible to zoom in the graphic view.
Use the Export results (CSV) button to get a results table useful for future processing.
Use the Export to image button to get the image of a reproducibility graph, useful for future documentation.