Report templates

Note

For a full explanation of how the report templates work, please see Report templating.

All analysis and comparison report templates are listed here. They are grouped by type (method/analysis and comparison) and ordered by name.

../../../_images/reportTemplates.png

For each report, you will see:

  • Type: in the method method, the analysis analyse or the comparison comparison group.

  • Name of the template.

  • Description.

  • Default: is the template the default template for its type (method, analysis or comparison). Each type has exactly one default template.

  • Built-in: is the template a standard visionCATS template (built-in report templates are read-only, and marked in blue).

  • Used: is the template currently in use. In this case, the report template cannot be deleted.

  • Legacy: is the template from before visionCATS 2.4 (legacy template cannot be used anymore).

Selecting a template will show which data is loaded with it in the report template options section. It is only modifiable for user template. The goal of this functionality is to reduce the loading time of the report by removing some data that are not displayed in the report. For example, the “Comparison profile overlay report” does not load data from the “Images” and “Spectrum” tab of the comparison.

For analysis report templates, the following loading options are available:

  • Load full plate images: whether to load full images, as displayed in the Data->Images tab

  • Load sample images: whether to load sample images, as displayed in Data->Samples tab

  • Load scanner track profiles: whether to load scanner 3/4 track profiles, as displayed in Data->Profiles tab

  • Load spectrum profiles: whether to load scanner 3/4 spectrum profiles, as displayed in Spectrum->Data tab

  • Load evaluation data: whether to load all evaluation data, including integrated profiles of the Evaluation-> Integration tab

  • Load log data: whether to load data of the HPTLC Logger

Note

Load log data is only available on visionCATS installations where the 21 CFR Part 11 Compliance option is active.

For comparison report templates, the options allow the user to choose whether to load data from the “Images”, “Profiles” and/or “Spectrum” tabs. By the way, selecting a report template will show the list of files where the template is used, in the Used in section. It is possible to open a file of the list by double-clicking on it.

Warning

Select only the needed data in your report templates, as more data is loaded, the more the report generation will take time.

From the toolbar, you can do the following:

  • duplicate duplicates selected template

  • download downloads all the files of the template (it is recommended to select an empty folder)

  • upload uploads all the files from selected folder, and will replace the selected template

  • delete deletes selected template

For non-built-in reports, the “Name” and “Description” fields are modifiable directly by clicking in the grid. Modifying report template contents consists in:

  1. Downloading the report template in a local folder

  2. Modifying the HTML/Javascript/CSS contents (see Changing template structure)

  3. Uploading the folder

  4. Clicking on the “Save” button.

To create a new report template:

  1. Select an existing one

  2. Hit “Duplicate”

  3. Click on the “Save” button.

To delete a report template (possible only for non-built-in and not used report templates):

  1. Select it,

  2. Hit “Delete”

  3. Click on the “Save” button.

Note

It is only possible to download legacy report templates: they are not usable anymore in analysis/comparison and visionCATS will display the default report template instead. If you want to “reuse” these templates, you should download them, open them in a Web browser (or any editor reading XML file), and then create a new template with a similar data structure.