DataView Toolbar
The data Toolbar contains all tools you can use on the image data.
Note
Depending on the size of the window, some tools may not appear directly. When available, use the bottom arrows next to each toolbar to use the hidden tools.
General tools
SelectThe default tool used to select or edit added elements on the images.
GrabGrab the image with this tool and move it in any direction to set a new view.
Export tool
Export imageExport image either to the file system or to the clipboard (to be pasted in another application (Word, Excel, etc…)). Export options allow to export the entire image or only the interesting area (tracks area), with or without annotations. In case of an export to the file system, the output DPI can be set. The exported image will also be cropped if Crop image is active.
𝑅ꜰ Tools and Tracks
Important
The 𝑅ꜰ tool/tracks positions should be correct by default, as defined in the Plate layout parameters. However, you may have to manually adjust it in some specific situations, for example when the plate wasn’t absolutely well positioned in the TLC Visualizer instrument.
Note
The color of the 𝑅ꜰ tool, tracks, annotations, tips and exposure normalization is set either to yellow, red or black, depending on the image illumination, in order appear clearly by default. You can change the color by right-clicking on the corresponding tool and select the appropriate contextual menu entry.
𝑅ꜰ ToolsToggle the visibility of the 𝑅ꜰ line and/or 𝑅ꜰ scale. Use the horizontal dotted lines to change the application and the front positions. Custom horizontal 𝑅ꜰ lines can be defined and saved to mark a specific position.
Reset 𝑅ꜰ Tool positionsReset the current 𝑅ꜰ Tool to its original position defined in the Plate layout parameters in Chromatography.
𝑅ꜰ TipMark a spot on the plate with the exact 𝑅ꜰ or metric value.
TracksToggle the visibility of the tracks to see where exactly does visionCATS “see” the tracks. Click on a track and move the vertical dotted lines in order to set the width of the tracks:
Adjust Tracks PositionUse the vertical dotted line on the left to set the position of the first track, and the vertical dotted line on the right to set the distance between tracks.
Reset positionReset the tracks position to its original values.
Export to comparisonExtract the selected track(s) and paste them into the comparison tool (see Export to comparison for more details)
Generate ProfilesGenerates the profiles of all the (non-clean) images of the current analysis based on the current position of the 𝑅ꜰ Tool.
Important
Deletes all output data (peaks, assignments, calibration functions and results) for evaluations based on a TLC Visualizer step. Only available if the current analysis does not contain any locked evaluation based on a TLC Visualizer step.
LabelsTurn automatic labeling for the tracks on or off. Select a combination of Vial ID, description, application volume or track number (i.e. track number) to be displayed.
Vial ID
Description
Application Volume (not applicable for tracks with overspots)
Track Number
Always hidden in the toolbar overflow:
𝑅ꜰ tool color
Track color
Labels text size
Image Settings
Important
Function Behavior:
Normalize Exposure: This function is exclusive, meaning that when it is enabled, all other settings or changes are reset. If any other function is applied afterward, the normalization itself will be reset.
SpotAmp, WhiteBalance, Clarify, and Color settings: These functions are cumulative, meaning that their effects can be applied together and will stack on top of one another.
Image settingsEnable or disable the following image editing features.
Note
If profiles have already been generated and the button is clicked, a popup asks for confirmation. This is because activating the settings will delete any existing profiles.
Spot AmpAmplify a region on the plate (increase contrast of an averaged small area on plate).
White BalanceSelect a small area on the image and adjust the color of the whole image to make this area exactly white.
Reset changesReset the Image settings adjustments (Spot Amp, Parameters and White Balance).
Crop imageRemove or show the 5mm background around each plate image.
Clean Plate CorrectionClean plate correction lets you subtract an image of the clean plate from images after chromatography. There is already some structure on the plate before chromatography (like background noise in an audio recording). The subtraction will result in a cleaner, sharpened image of the plate.
Important
To do the subtraction, the plate should be on the same position on both images. visionCATS will try to detect any shift or rotation, but in some cases (especially for derivatized step) no matching is found. It is possible to still use the correction if wanted, but a visual check should be done to verify that no big shift or rotation exists (otherwise the correction would do more harm than good).
If already written on clean plate, the plate identifier notation makes a good indicator of the correction quality:
A perfect clean plate notation will make the text practically invisible, without any shadow effect.
A small shift will create some shadow on text, but the correction should still be beneficial.
A big shift will be more like two layers of text, it is recommended to disable clean plate correction in this case.
Warning
CPU missing AVX or AVX2 instruction will lead to visionCATS’ client crash when using this feature.
HDRI Settings
The following tools are only available for HDRI images (which are only available for R366).
Normalize exposureThe saturation of the defined zone (exposure range) is normalized to 85%. Only a standard-dynamic-range around the wanted exposure will be displayed, all the data outside the range will be overexposed (white) or underexposed (black). If the exposure range is placed on a representative track of the plate (normally a reference), the important information of the image will be correctly displayed.
Note
Enabling normalization will disable Clarify.
Show exposure rangeThe current exposure range is displayed and can be modified graphically.
Reset exposure rangeThe exposure range is reset to the original value, which is the first track of the plate.
ClarifyVirtually change the illumination setting after capturing. Makes weak spots visible. It is a tone mapping technique which will compress the dynamic-range to makes most information visible on a limited dynamic-range monitor. The compression result in different color saturation and contrast than the reality, but more information is visible.
Enhancement slider: how much compression of dynamic-range will be done, making more or less information visible
Annotations
TextCreate a text box to put a comment on the plate image.
ArrowIndicate a specific area on the plate image with an arrow.
DeleteDelete the selected Annotation items.
CopyCopy the selected Annotation items.
PastePaste the selected Annotation items.
Always hidden in the toolbar overflow:
ColorChange the color of selected Annotation items.
BoldSet text weight of selected Annotation items to bold.
ItalicSet text weight of selected Annotation items to italic.
UnderlineSet text weight of selected Annotation items to underline.
Rotate LeftRotate selected Annotation items counter clock-wise.
Right RotateRotate selected Annotation items clock-wise.
FontSet text font of selected Annotation items.
Text sizeSet text size of selected Annotation items.